1 transmitter output connections (section #1), Comment – CiDRA SONARtrac HD VF-100 User Manual
Page 45
Copyright © 2006 CiDRA Corporation
Page 8-5
20638-02 Rev 01
8.4.2.1 Transmitter
Output Connections (Section #1)
The following figure provides a close-up of the transmitter output
terminals (Section #1 of the terminal board) with their functions listed
in the following table. These outputs can be connected as appropriate
to permit communications between the transmitter and other
equipment.
Figure 36
Transmitter Output Terminals
Terminal Label
Type
Comment
CUR 1
4-20mA #1 Output
Internal (self) or external (loop) powered,
HART communication capable
CUR 2
4-20mA #2 Output
Internal (self) or external (loop) powered
PULSE
Pulse output
Solid-state relay closure
ALARM
Alarm output – High /
Low Output
Solid-state relay closure
COMM
RS-485 or 232 Output
User configured
SHD ---
Shield
Connections
Table 5
List of Transmitter Signal Outputs
User supplied data output cable (size range AWG 30 to AWG 10) is
installed through a cable gland in the furthest left hole on the
transmitter housing and attached to the appropriate terminal block
connection points. The cable gland is sized for a 3/4 inch NPT fitting
(1-1/16 inch hole).
WARNING
For Class I, Division 2 applications, Power Entry and
Inputs/Outputs must be installed in accordance with Article
501.10(B)(1) of the National Electrical Code ANSI/NFPA 70:2005