6400/6404 Series Operations Manual

Page 50 of 54

Rev F 09/18/14

SECTION 12 – PREVENTATIVE MAINTENANCE

Your CARON incubator is robustly designed to minimize performance problems.

However, regular maintenance is very important for continuous trouble free operation.

As a general rule, CARON recommends an annual calibration check of the temperature,

humidity (optional), and CO2 systems. CARON offers a full range of on-site calibration

and validation services. We also offer preventative maintenance contracts on our

equipment. Contact our customer service department for details or visit us on the web

at

www.caronproducts.com

.

Recommended Daily Maintenance Checks

Check the Temperature and CO2 displays versus set-points.

Check for and correct any alarm condition.

Check the CO2 and N2 gas tank levels

Check the water level in the humidity pan

Recommended Annual Maintenance Checks

Replace inline HEPA filters on both CO2 and N2 gas supplies

Replace O2 sensor

Check tightness of hinge screws on inner and outer door moving parts

Disinfect all interior surfaces with a general purpose laboratory cleaning agent.

Perform a complete calibration of the temperature, humidity, and CO2, and O2

systems.

A full validation is recommended for GMP facilities each time a unit is installed,

moved or undergoes significant repair. Contact CARON’s service department to

schedule on-site validation.