6400/6404 Series Operations Manual
Page 50 of 54
Rev F 09/18/14
SECTION 12 – PREVENTATIVE MAINTENANCE
Your CARON incubator is robustly designed to minimize performance problems.
However, regular maintenance is very important for continuous trouble free operation.
As a general rule, CARON recommends an annual calibration check of the temperature,
humidity (optional), and CO2 systems. CARON offers a full range of on-site calibration
and validation services. We also offer preventative maintenance contracts on our
equipment. Contact our customer service department for details or visit us on the web
Recommended Daily Maintenance Checks
Check the Temperature and CO2 displays versus set-points.
Check for and correct any alarm condition.
Check the CO2 and N2 gas tank levels
Check the water level in the humidity pan
Recommended Annual Maintenance Checks
Replace inline HEPA filters on both CO2 and N2 gas supplies
Replace O2 sensor
Check tightness of hinge screws on inner and outer door moving parts
Disinfect all interior surfaces with a general purpose laboratory cleaning agent.
Perform a complete calibration of the temperature, humidity, and CO2, and O2
systems.
A full validation is recommended for GMP facilities each time a unit is installed,
moved or undergoes significant repair. Contact CARON’s service department to
schedule on-site validation.