6400/6404 Series Operations Manual

Page 27 of 54

Rev F 09/18/14

Calibrating the Relative Humidity Display

If your incubator is equipped with an optional humidity sensor and display, the humidity

system can be calibrated. If the humidity option was not purchased, the calibration

menus will not be displayed in the calibration mode. Place the relative humidity

reference device to which the incubator temperature will be calibrated to in the center of

the chamber. If the incubator has just been installed and started up, allow 8-12 hours for

the incubator to stabilize before making a calibration adjustment. If the unit has been in

operation, allow a full hour after the temperature display reaches the temperature

setpoint before making any calibration adjustments.

1) Press the key from the main menu

2) Press the to select Calibration

3) Press the key to enter Calibration

4) Press the to select Rh Sensor

5) Press the key to enter Rh Sensor

6) Press the to increase / decrease the Rh Sensor value (hold for scroll)

7) Press the key to save the Rh Sensor calibration value

8) Press the key twice to return to the main menu