6400/6404 Series Operations Manual
Page 33 of 54
Rev F 09/18/14
Setting High/Low O2 Alarm Level (suppressed O2 model 6404s only)
The incubator has both a high and low programmable O2 alarm. The factory default for
the high O2 alarm is 1.0% above O2 setpoint. The factory default for the low
Temperature alarm is 1.0% below O2 setpoint. To change either alarm level:
1) Press the key from the main menu
2) Press the to select Admin
3) Press the key to enter Admin
4) Press the to select High O2 Alarm / Low O2 Alarm
5) Press the key to enter High O2 Alarm / Low O2 Alarm
6) Press the to change alarm level
7) Press the key to save the alarm level
8) Press the key twice to return to the main menu