Caron 6404-4 User Manual

Page 33

background image

6400/6404 Series Operations Manual

Page 33 of 54

Rev F 09/18/14

Setting High/Low O2 Alarm Level (suppressed O2 model 6404s only)

The incubator has both a high and low programmable O2 alarm. The factory default for

the high O2 alarm is 1.0% above O2 setpoint. The factory default for the low

Temperature alarm is 1.0% below O2 setpoint. To change either alarm level:

1) Press the key from the main menu

2) Press the to select Admin

3) Press the key to enter Admin

4) Press the to select High O2 Alarm / Low O2 Alarm

5) Press the key to enter High O2 Alarm / Low O2 Alarm

6) Press the to change alarm level

7) Press the key to save the alarm level

8) Press the key twice to return to the main menu