Caron 6404-4 User Manual
Page 14

6400/6404 Series Operations Manual
Page 14 of 54
Rev F 09/18/14
Leveling the Unit
Place a level on the middle shelf of the incubator. The cabinet has built in adjustable
leveling feet designed to level the unit. Adjust the feet appropriately until the unit sits
level left to right and front to back.
Installing Filtered Air Exchange Port Stopper
The Oasis incubator is designed with a filtered air exchange that allows a small amount
of ambient air to be drawn into the incubator to slightly reduce the relative humidity to
approximately 90% and just below saturation. The recirculating fan pulls HEPA filtered
ambient air into the incubator and exhausts the air through a HEPA filtered stopper. The
silicone port stopper HEPA assembly that is supplied in the accessory kit will need to be
installed in the rear access port located on the top left side of the back wall as seen in
the picture below.
If higher humidity levels higher than 90% are required, install the brass plug that is
supplied in the accessory kit into the rear panel of the cabinet labeled Control Leak.
This will disable the filtered air exchange. With the filtered air exchange disabled, the
humidity level may increase above saturation resulting in some condensation developing
on incubator surfaces.
Before relocating an incubator remove the water
in the humidity pan and transfer any product to
another incubator.