6400/6404 Series Operations Manual

Page 14 of 54

Rev F 09/18/14

Leveling the Unit

Place a level on the middle shelf of the incubator. The cabinet has built in adjustable

leveling feet designed to level the unit. Adjust the feet appropriately until the unit sits

level left to right and front to back.

Installing Filtered Air Exchange Port Stopper

The Oasis incubator is designed with a filtered air exchange that allows a small amount

of ambient air to be drawn into the incubator to slightly reduce the relative humidity to

approximately 90% and just below saturation. The recirculating fan pulls HEPA filtered

ambient air into the incubator and exhausts the air through a HEPA filtered stopper. The

silicone port stopper HEPA assembly that is supplied in the accessory kit will need to be

installed in the rear access port located on the top left side of the back wall as seen in

the picture below.

If higher humidity levels higher than 90% are required, install the brass plug that is

supplied in the accessory kit into the rear panel of the cabinet labeled Control Leak.

This will disable the filtered air exchange. With the filtered air exchange disabled, the

humidity level may increase above saturation resulting in some condensation developing

on incubator surfaces.

Before relocating an incubator remove the water

in the humidity pan and transfer any product to

another incubator.