6400/6404 Series Operations Manual

Page 30 of 54

Rev F 09/18/14

SECTION 7 – ADMIN MODE

Many incubator settings are set to common defaults to minimize initial setup of the

incubator. However, these settings are also adjustable to allow a user to customize the

incubator. These adjustments are all available in ADMIN Mode.

Setting Over/Low Temperature Alarm Level

The incubator has both an over and low programmable temperature alarm. The factory

default for the over temperature alarm is 1.0°C above setpoint. The factory default for

the low temperature alarm is 1.0°C below setpoint. To change either alarm level:

1) Press the key from the main menu

2) Press the to select Admin

3) Press the key to enter Admin

4) Press the to select Over Temp Alarm / Low Temp Alarm

5) Press the key to enter Over Temp Alarm / Low Temp Alarm

6) Press the to change alarm level

7) Press the key to save the alarm level

8) Press the key twice to return to the main menu