Caron 6404-4 User Manual
Page 30

6400/6404 Series Operations Manual
Page 30 of 54
Rev F 09/18/14
SECTION 7 – ADMIN MODE
Many incubator settings are set to common defaults to minimize initial setup of the
incubator. However, these settings are also adjustable to allow a user to customize the
incubator. These adjustments are all available in ADMIN Mode.
Setting Over/Low Temperature Alarm Level
The incubator has both an over and low programmable temperature alarm. The factory
default for the over temperature alarm is 1.0°C above setpoint. The factory default for
the low temperature alarm is 1.0°C below setpoint. To change either alarm level:
1) Press the key from the main menu
2) Press the to select Admin
3) Press the key to enter Admin
4) Press the to select Over Temp Alarm / Low Temp Alarm
5) Press the key to enter Over Temp Alarm / Low Temp Alarm
6) Press the to change alarm level
7) Press the key to save the alarm level
8) Press the key twice to return to the main menu