The decon cycle is now initiated. Do not open the outer door during the
decontamination process. The message center on the display will notify you how much
time remains to complete the cycle and when the cycle is complete.
The total time to complete the decon cycle will vary slightly depending upon ambient
conditions and line voltage to the incubator. The overall time to complete the cycle is
approximately 15 hours. Heat up time is approximately 2 hours followed by 9 hours of
soak time at elevated temperature and humidity. The drying cycle lasts 30 minutes,
followed by a cool down time of approximately 3 hours.
Upon completion of the decon cycle, fill the humidity pan as instructed in the installation
instructions. Replace the IR sensor reversing the process followed to remove it.
Replace the O2 sensor reversing the process followed to remove it. Press ENTER for
the incubator to return to normal operation mode.