Changing the Temperature Set-point

The incubator has an operating temperature range of 20.0°C to 60.0°C. The lowest

temperature set-point that the incubator will control temperature is 5°C above ambient.

A setting of 20.0°C will disable the temperature control and all heaters and alarms will

be disabled. To change the operating temperature setpoint:

1) Press the key from the main menu

2) Press the to select settings

3) Press the key to enter Settings

4) Press the to select Set Temp

5) Press the key to enter Set Temp

6) Press the to increase / decrease temperature setpoint (hold for scroll)

7) Press the key to save the setpoint

8) Press the key twice to return to the main menu