6400/6404 Series Operations Manual
Page 39 of 54
Rev F 09/18/14
Configuring the Temp / CO2 / RH / O2 Alarm Relay
The incubator is equipped a set of dry alarm contacts. Rating of the relay contacts is a
maximum of 30V DC, 2A. A set of NO (normally open) and NC (normally closed)
contacts are available on the back of the incubator. The software can configure the
relay so it changes state during a temperature or CO2 alarm. To enable/disable a temp
/ CO2 / RH / O2 alarm from changing the state of the relay:
1) Press the key from the main menu
2) Press the to select Admin
3) Press the key to enter Admin
4) Press the to select the desired alarm relay configuration
5) Press the key to modify the desired alarm relay configuration
6) Press the to toggle between Disable / Enable
7) Press the key to save the alarm relay configuration
8) Press the key twice to return to the main menu