SECTION 10 – ALARMS

A sophisticated alarm system monitors all system parameters for any fault condition. At

the same time, built in alarm delays protect the user from nuisance alarms created by

routine use such as door openings. A list of alarm descriptions, the conditions they look

for, and how they are delayed and displayed are in the table below:

Alarm Description

Condition

Detected

Delay*

Audible

/ Visual

or Both

Alarm

Rings

Back

Other Notes

TEMP SENSOR ERROR Faulty Temp Sensor

Detected

No Delay

Both

Yes

Display shows “---“; Heating will

be disabled

CO2 SENSOR ERROR

Faulty CO2 Sensor

Detected

No Delay

Both

Yes

Display shows “---“; CO2 will be

disabled

RH SENSOR ERROR

Faulty Humidity

Sensor Detected

No Delay

Both

Yes

Display shows “---“;

OVER TEMPERATURE

Temperature reading

greater than

Overtemp Alarm

No Delay

Both

Yes

Heating is disabled, Display

continues to show Temp

reading

HIGH CO2 LEVEL

CO2 reading greater

than Overtemp

Alarm

15 Minute Delay of

Continuous High Reading

Both

Yes

CO2 injection is disabled,

display continues to show CO2

reading

DOOR OPEN

Outer Door Open

Programmable 1-15

minutes

Both

Yes

CO2 injection is disabled,

internal fan shuts down

LOW % RH

(optional)

Humidity reading

Lower than Low

%RH Alarm

Startup Delay** & 30

Minute Delay of

Continuous Low Reading

Both

Yes

Display continues to show Rh

reading

LOW TEMPERATURE

Temperature reading

lower than Low

Temp Alarm

Startup Delay** & 15

Minute Delay of

Continuous Low Reading

Both

Yes

Display continues to show

Temp reading

LOW CO2 LEVEL

CO2 reading lower

than Low CO2 Alarm

Startup Delay** & 15

Minute Delay of

Continuous Low Reading

Both

Yes

Display continues to show CO2

reading

HIGH O2 LEVEL

O2 reading greater

than High O2 alarm

Startup Delay** & 15

Minute Delay of

Continuous High Reading

Both

Yes

Display continues to show O2

reading

LOW O2 LEVEL

O2 reading lower

than Low O2 Alarm

15 Minute Delay of

Continuous Low Reading

Both

Yes

N2 injection is disabled, display

continues to show O2 reading

ALL CO2 TANKS LOW

(optional)

Optional Gas Guard

indicates both CO2

tanks are low

No Delay

Both

No

Display continues to show CO2

reading

CO2 TANK 1 LOW

(optional)

Optional Gas Guard

indicates tank 1 has

low pressure

15 Minute Delay of

Continuous Low Reading

Both

No

Solenoid switches to tank 2

inlet. Display continues to

show CO2 reading

CO2 TANK 2 LOW

(optional)

Optional Gas Guard

indicates tank 2 has

low pressure

15 Minute Delay of

Continuous Low Reading

Both

No

Solenoid switches to tank 1

inlet. Display continues to

show CO2 reading

ALL N2 TANKS LOW

(optional)

Optional Gas Guard

indicates both N2

tanks are low

No Delay

Both

No

Display continues to show O2

reading

N2 TANK 1 LOW

(optional)

Optional Gas Guard

indicates N2 tank 1

has low pressure

15 Minute Delay of

Continuous Low Reading

Both

No

Solenoid switches to tank 2

inlet. Display continues to

show O2 reading

N2 TANK 2 LOW

(optional)

Optional Gas Guard

indicates N2 tank 2

has low pressure

15 Minute Delay of

Continuous Low Reading

Both

No

Solenoid switches to tank 1

inlet. Display continues to

show O2 reading

* Status message is shown in the message center status menu immediately; unless in startup delay.

** To avoid nuisance alarms on startup, alarm is delayed until parameter reaches alarm limit the first time.