6400/6404 Series Operations Manual

Page 19 of 54

Rev F 09/18/14

Accessory FLST402 in the figure below provides swivel casters for added mobility to

either single or stacked incubators. The two front casters have brakes which can be

locked once the unit is rolled to its final location. The unit is designed for a single

incubator and elevates it from the floor by approximately 25”.

Connecting the CO2 or N2 Gas Guard

Accessory GASG401 CO2 backup system and GASG403 N2 backup system provide an

optional built in system to allow two regulated CO2 or N2 supplies to connect to an

incubator.

Refer to the equipment installation section, connecting a CO2 or N2 gas supply.

The gas guard is designed to automatically switch from the primary tank to the

secondary tank when low gas pressure is detected on the primary tank. This allows for

a temporary uninterrupted supply of CO2 or N2 to the incubator should the primary tank

go empty.