Setting High/Low CO2 Alarm Level
The incubator has both a high and low programmable CO2 alarm. The factory default
for the high CO2 alarm is 1.0% above CO2 setpoint. The factory default for the low
Temperature alarm is 1.0% below CO2 setpoint. To change either alarm level:
1) Press the key from the main menu
2) Press the to select Admin
3) Press the key to enter Admin
4) Press the to select High CO2 Alarm / Low CO2 Alarm
5) Press the key to enter High CO2 Alarm / Low CO2 Alarm
6) Press the to change alarm level
7) Press the key to save the alarm level
8) Press the key twice to return to the main menu