Changing the O2 Set-point
The incubator has an operating O2 range of 20.6% O2 to 1.0% O2. A setting of either
20.7% or 1.0% O2 will disable the O2 control and all O2 control and alarms. To change
the operating O2 setpoint:
1) Press the key from the main menu
2) Press the to select settings
3) Press the key to enter Settings
4) Press the to select Set O2
5) Press the key to enter Set O2
6) Press the to increase / decrease O2 setpoint (hold for scroll)
7) Press the key to save the setpoint
8) Press the key twice to return to the main menu