Caron 6404-4 User Manual
Page 17

6400/6404 Series Operations Manual
Page 17 of 54
Rev F 09/18/14
Connect each incubator to a dedicated
grounded circuit. Failure to do so could result in
electrical shock. Ensure that there is
unobstructed access to the main power plug
disconnect.
Connecting Electrical Power
Model 6400-1 requires a 100-130 VAC, 50/60Hz 15A power connection. The power
cord connection is a NEMA 5-15P plug.
Model 6400-4 requires a 220 – 240 VAC, 50/60Hz, 16A power connection. The power
cord connection is a CEE 7/7 plug.
When the required electrical connection is available, plug the provided power cord into
the unit and the electrical outlet.
Connecting Alarm Contacts
A set of dry relay alarm contacts are provided standard with each incubator. They are
located on the back of the incubator. The terminals on the relay contact allow for a NO
(normally open) output, a NC (normally closed) and COM (common) connection.
In the event of an alarm condition or power failure, the NO contact will close and the NC
contact will open. Once the alarm is cleared, the contacts will return to their normal
position.