Connect each incubator to a dedicated

grounded circuit. Failure to do so could result in

electrical shock. Ensure that there is

unobstructed access to the main power plug

disconnect.

Connecting Electrical Power

Model 6400-1 requires a 100-130 VAC, 50/60Hz 15A power connection. The power

cord connection is a NEMA 5-15P plug.

Model 6400-4 requires a 220 – 240 VAC, 50/60Hz, 16A power connection. The power

cord connection is a CEE 7/7 plug.

When the required electrical connection is available, plug the provided power cord into

the unit and the electrical outlet.

Connecting Alarm Contacts

A set of dry relay alarm contacts are provided standard with each incubator. They are

located on the back of the incubator. The terminals on the relay contact allow for a NO

(normally open) output, a NC (normally closed) and COM (common) connection.

In the event of an alarm condition or power failure, the NO contact will close and the NC

contact will open. Once the alarm is cleared, the contacts will return to their normal

position.