Caron 6404-4 User Manual
Page 37
6400/6404 Series Operations Manual
Page 37 of 54
Rev F 09/18/14
Setting the Alarm Ring Back Time
The incubator is equipped with an audible alarm ring back timer. The ring back timer will
begin when an alarm is silenced using the silence key on the control panel. When the
ring back timer hits the programmed value, the audible alarm will sound again,
reminding the user that the alarm condition needs addressed. The factory default for the
alarm ring back timer is 15 minutes. To change the alarm ring back timer:
1) Press the key from the main menu
2) Press the to select Admin
3) Press the key to enter Admin
4) Press the to select Alarm Ring Back
5) Press the key to enter Alarm Ring Back
6) Press the to adjust the Alarm Ring Back Setting
7) Press the key to save the Alarm Ring Back setting
8) Press the key twice to return to the main menu