6400/6404 Series Operations Manual

Page 45 of 54

Rev F 09/18/14

Use only distilled or deionized water with a

resistivity between 50K

-CM and 1M

-CM and

neutral pH. Using water outside this range will

void your warranty.

SECTION 9 – USING THE AUTOMATIC DECON CYCLE

The incubator is equipped with a built-in automatic 90°C moist heat decontamination

cycle. The cycle is designed to decontaminate the entire interior workspace including all

installed components. Moist heat at 90°C is proven to effectively kill common organisms

known to contaminate incubators.

The decon cycle consists of a heating phase in which the temperature and humidity rise

to approximately 90°C and 60% relative humidity. After reaching these temperatures

the unit soaks for several hours before entering a drying phase. During the drying

phase the heaters maintain 90°C while a built in pump pulls in HEPA filtered clean air.

Next, the unit cools down to the setpoint from which the decon cycle was initiated. The

total cycle time is approximately 15 hours. The graph below represents the various

cycles of the decon cycle:

To initiate the decon cycle press the DECON button on the control panel and follow the

on screen prompts.

The first step of the cycle is to replace the water in the pan with 400 mL of water. This is

the amount of water required to effectively create the moist heat decontamination

environment.