6400/6404 Series Operations Manual

Page 20 of 54

Rev F 09/18/14

Connecting Analog Outputs

Accessory OUTP401 provides terminals at the rear of the incubator to output analog

signals, representing the temperature, CO2, and optional relative humidity levels inside

the incubator. Connections can be made to in house data acquisition systems to

monitor incubator performance. The accessory can be configured at the factory to

supply either 4-20mA or 0-5V signals that represent the parameters.

The table below represents the range of parameter values

Controller Value

Output 0-5V

Output 4-20mA

Temperature Low

0 degrees C

0.0 VDC

4 mA

Temperature High

100 degrees C

5.0 VDC

20 mA

CO2 Low

0.0% CO2

0.0 VDC

4 mA

CO2 High

20.0% CO2

5.0 VDC

20 mA

Humidity Low

0% rH

0.0 VDC

4 mA

Humidity High

100% rH

5.0 VDC

20 mA

O2 Low

0.0% O2

0.0 VDC

4 mA

O2 High

20.7% O2

5.0 VDC

20 mA