Connecting Analog Outputs
Accessory OUTP401 provides terminals at the rear of the incubator to output analog
signals, representing the temperature, CO2, and optional relative humidity levels inside
the incubator. Connections can be made to in house data acquisition systems to
monitor incubator performance. The accessory can be configured at the factory to
supply either 4-20mA or 0-5V signals that represent the parameters.
The table below represents the range of parameter values
Controller Value
Output 0-5V
Output 4-20mA
Temperature Low
0 degrees C
0.0 VDC
4 mA
Temperature High
100 degrees C
5.0 VDC
20 mA
CO2 Low
0.0% CO2
0.0 VDC
4 mA
CO2 High
20.0% CO2
5.0 VDC
20 mA
Humidity Low
0% rH
0.0 VDC
4 mA
Humidity High
100% rH
5.0 VDC
20 mA
O2 Low
0.0% O2
0.0 VDC
4 mA
O2 High
20.7% O2
5.0 VDC
20 mA