Caron 6404-4 User Manual
Page 23

6400/6404 Series Operations Manual
Page 23 of 54
Rev F 09/18/14
Changing the CO2 Set-point
The incubator has an operating CO2 range of 0.0% CO2 to 20.0% CO2. A setting of
0.0% CO2 will disable the CO2 control and all CO2 control and alarms. To change the
operating CO2 setpoint:
1) Press the key from the main menu
2) Press the to select settings
3) Press the key to enter Settings
4) Press the to select Set CO2
5) Press the key to enter Set CO2
6) Press the to increase / decrease CO2 setpoint (hold for scroll)
7) Press the key to save the setpoint
8) Press the key twice to return to the main menu