Troubleshooting
Eppendorf

®

PlateReader AF2200

English (EN)

74

7.3

Troubleshooting Fluorescence mode

7.4

Save method scripts

Predefined methods, with the exception of Nucleic acid quantification (UV 260 nm microvolume), may be
modified by the user and saved under another name (Save-as functionality) in the directory:

• Win XP: C:\ Documents and Settings\All Users\Documents\Eppendorf\Methods

• Win 7: C:\Users\Public\Documents\Eppendorf\Methods

These methods can be loaded for further measurements by selecting

Open from method

from the File menu.

7.5

Save method results

After performing a measurement using a predefined method, the results are automatically saved in the
following directory:

• Win XP: C:\ Documents and Settings\All Users\Documents\Eppendorf\Results

• Win 7: C:\Users\Public\Documents\Eppendorf\Results

Warning / error

Consequences

Remedy

Sample data out of
min-max-range of standard
values.

Corresponding sample results are
marked red in the results table.

Dilute sample or modify standard
concentrations and remeasure.

Correlation coefficient of standard
curve < user input.

Sample results are not calculated. Check input of standard

concentrations and remeasure.
Check concentrations of standard
solutions and remeasure
Reduce your required value of R

2

in the

User Settings

Dialog of the

software and remeasure.

At least one sample is out of
measurement range.

Raw data of the samples are
displayed as

OVER

, sample results

are not calculated.

Check gain settings and
remeasure.
Dilute sample and remeasure.

At least one sample is out of
measurement range.

Raw data are displayed as

OVER

,

standard curve is not calculated.

Check gain settings and
remeasure.
Dilute standard and remeasure.

Calculation Error

Standard curve is not calculated

Define standard concentrations in
the method stripe of the software
and remeasure.