4 kinetic – Eppendorf AF2200 Plate Reader User Manual

Page 20

background image

Overview
Eppendorf

®

PlateReader AF2200

English (EN)

20

Shaking

1. Select the Shaking program element if the plate is to be shaken, either before the measurement or

between kinetic cycles.

2. Enter the respective parameters (see Parameter list).

3. Clicking the link

Wait a couple of seconds

inserts a new program element (see Miscellaneous on p. 21).

Tab. 4-1: Parameter list

4.2.4

Kinetic

Kinetic Cycle

Use the program element Kinetic Cycle to perform several consecutive measurements, which may be
executed in certain intervals.

1. Enter the respective parameters:

Plate-wise kinetic measurements

Each cycle of the kinetic measurement is performed on all selected wells. Plate-wise kinetic measurements
may contain a maximum of ten independent measurement stripes that do not need to be of the same
measurement type.

Duration

Enter the duration of the shaking process.

Mode

Select between the options

Linear

and

Orbital

.

Amplitude

Enter the required Amplitude value from the drop-down list.

Cycles

Number of cycles:

Enter a number or click the up or down arrows for

the number of actual measurement steps (2 – 1000 cycles)

Duration:

Enter the duration, format hh:mm:ss.

Kinetic Interval

Use kinetic interval:

Enter the time interval (hh:mm:ss or ms).