Overview

Eppendorf

®

PlateReader AF2200

English (EN)

21

Well-wise kinetic measurements

All cycles of the kinetic measurement are first performed in one well before continuing to the next well.
Well-wise kinetic measurements may be composed of a maximum of four measurement stripes of the same
type, e.g., four absorbance stripes.

Kinetic Condition

Use the Kinetic Condition program element to define which actions should be executed at a certain cycle.

If 3 is entered for

Execute command at cycle

within a kinetic measurement containing, e.g. a Shake step,

shaking is performed only at cycle 3.

4.2.5

Miscellaneous

Comment

Use the program element Comment to enter a remark or statement for the current measurement in the text
field. This text is shown together with the measurement in the Excel output sheet.

Kinetic conditions should be inserted right after a Kinetic Cycle program element in order to
ensure optimal result reproducibility.
Users are advised to set up suitable scripts prior to the measurements and to use the same
script for all similar kinetic measurements in order to obtain comparable results.