Troubleshooting

Eppendorf

®

PlateReader AF2200

English (EN)

7

Troubleshooting

7.1

General Troubleshooting

If you don’t meet your expected measurement results please check the following:

• The volumes of all measured wells are homogeneous. Uniformity of pathlength [d] is essential for

reliable measurements.

• The homogeneity of your samples is given:

The bottom of the wells must be entirely covered with fluid.

Precipitation and irregular colour product distribution have to be avoided.

Buffers which have massive meniscus effects shouldn’t be used.

7.2

Troubleshooting Absorbance mode

Identical volumes of different fluid types and buffers may have different pathlengths.

Warning / error

Consequences

Remedy

Sample data out of
min-max-range of standard values

Corresponding sample results are
marked red in the results table.

Dilute sample or modify standard
concentrations and remeasure.

Correlation coefficient < 
user input

Sample results are not calculated. Check input of standard

concentrations and remeasure.
Check concentrations of standard
solutions and remeasure
Reduce your required value of R

2

in the

User Settings

Dialog of the

software and remeasure.

At least one sample is out of
measurement range.

Raw data of the samples are
displayed as

OVER

, sample results

are not calculated.

Dilute sample and remeasure.

At least one standard is out of
measurement range.

Raw data are displayed as

OVER

,

standard curve is not calculated.

Dilute standard solution and
remeasure.