Predefined methods
Eppendorf

®

PlateReader AF2200

English (EN)

62

User defined method parameters

• Using the predefined UV/Vis filter slide the measurement is performed at 600 nm with default

measurement parameters.

• Define the number of blank replicates (possible number: 1 – 8).

• Define the number of standards (possible number: 2 – 12).

• Define number of standard replicates (possible number: 1 – 8).

• Define the concentrations of each standard (Press triangle button to open the menu).

• Define the number of samples (max. 93, depends on number of blanks and standards).

• Define the unit of your concentration.

• Check if entered parameters are correct.

• Required plate layout: Pipette in a vertical order, starting with blanks at position A1. Standards are

pipetted in subsequent wells, samples must be placed in the following wells.

• Press

Start

.

• For method results and data evaluation principle see chapter

method results

and

evaluation procedure

.

Plate layout

Abb. 6-5: Method strip: Protein quantification (BCA 562 (600) nm with standards)

Fig. 6-5:

Method strip: Protein quantification (BCA 562 (600) nm with standards)

The selected number of blanks, standards and samples are shown in the plate view.