Predefined methods

Eppendorf

®

PlateReader AF2200

English (EN)

57

User defined method parameters

• The measurement is performed at 260 nm with default measurement parameters.

• Define the number of blank replicates (possible number: 1 – 8).

• Define the number of standards (possible number: 2 – 12).

• Define number of standard replicates (possible number: 1 – 8).

• Define the concentrations of each standard (Press triangle button to open the menu).

• Define the number of samples (max. 93, depends on number of blanks and standards).

• Define the unit of your concentration.

• Check if entered parameters are correct.

• Required plate layout: Pipette in a vertical order, starting with blanks at position A1. Standards are

pipetted in subsequent wells, samples must be placed in the following wells.

• Press

Start

.

• For method results and data evaluation principle see chapter

method results

and

evaluation procedure