2 error messages, 1 warnings, 2 error – Eppendorf TransferMan4 m User Manual
Page 79: Error messages 7.2.1, Warnings, Error

Troubleshooting
TransferMan
®
4m
English (EN)
79
7.2
Error messages
7.2.1
Warnings
7.2.2
Error
Symptom/message
Cause
Remedy
WARNING 1
• FemtoJet is not connected.
Disconnect the data cable to
the FemtoJet and connect it
again.
Switch on the FemtoJet.
WARNING 3
• X-module is not connected.
Switch off the device.
Connect the plug of the
X-module to the control
board and tighten.
Switch the device on.
WARNING 4
• Y-module is not connected.
Switch off the device.
Connect the plug of the
Y-module to the control
board and tighten.
Switch the device on.
WARNING 5
• Z-module is not connected.
Switch off the device.
Connect the plug of the
Z-module to the control
board and tighten.
Switch the device on.
WARNING 6
• Synchronization error with
FemtoJet during an
injection.
Set another synchronization
for
Synchr. inject.
.
Set the synchronization
IMMEDIATE
,
LIMIT
or
PRESSURE
.
Symptom/message
Cause
Remedy
ERROR 10
–
ERROR 99
• Technical error.
Switch the device off and
back on.
Reset the parameters to the
factory settings.
Contact Eppendorf Service.