1 manually positioning the capillary, 2 automatically positioning the capillary, Manually positioning the capillary – Eppendorf TransferMan4 m User Manual
Page 55: Automatically positioning the capillary
Operation
TransferMan
®
4m
English (EN)
55
1. Move the capillary out of the working range using the
home
key.
2. Swing the X-module forwards.
3. Undo the grip head on the capillary holder.
4. Carefully pull the capillary out of the grip head.
5. Push the new capillary into the grip head until it reaches the stop and tighten the grip
head.
6. Swing the X-module back.
6.4.1
Manually positioning the capillary
1. Press
Back manual
.
2. Manually position the capillary in the working range.
6.4.2
Automatically positioning the capillary
1. Press
home
.
• The capillary automatically moves back into the working range.
CAUTION! Risk of crushing between the modules
The modules automatically move in all spatial axes.
Do not reach into the movement range of the modules.
Suitable when using capillaries of different lengths (e.g., self-pulled capillary).
Suitable when using industrial capillaries (e.g., Femtotips).