41

Prerequisites

• TransferMan 4m is switched off.

• The power cable is disconnected.

1. Connect the module (X,Y,Z) plug with the ports on the control board.

2. Tighten the fixing screws on the plug manually.

3. Connect the power cable.

4. Switch on the mains switch.

5. Set the installation parameters. You can use the software wizard

First set-up

or the

Installation

menu to set the

Side

and

Angle

parameters.

4.13

Setting installation parameters

Installation parameters must be set:

• for the initial installation

• after a reset

The following settings are defined:

• Mounting side of the motor module

• Operating angle for the capillary

• Center motors

• Adjust motors

• Set date

• Dynamic movement range of the joystick

NOTICE! Material damage from incorrect connections.

Only electrical connections may be made to devices described in the
operating manual.

Other connections are permitted only following consultation and agreement
with Eppendorf AG.

Only connect devices that meet the safety requirements defined in
IEC 60950-1.

NOTICE! Short circuit caused by incorrect installation.

Failure to observe the order of steps may result in a short circuit.