beautypg.com

2 executing the step injection function, 13 home function, 1 home menu and parameters – Eppendorf TransferMan4 m User Manual

Page 64: Executing the step injection function, Home function 6.13.1, Home menu and parameters

background image

Operation
TransferMan

®

4m

English (EN)

64

6.12.2

Executing the

Step injection

function

Prerequisites

• A FemtoJet is connected.

Step injection

is assigned to a free softkey.

1. Activate

Step injection

.

2. Set

Step inj. dist.

3. Set the injection time on the FemtoJet.

4. Select

Step inj. dist.

5. Close the menu.

• The application screen displays

Step injection

.

6. Trigger the injection by selecting

Inject

on the FemtoJet.

• The function is executed.

6.13

Home

function

The

Home

function quickly moves the capillary out of the work area and is suitable for a

quick change of capillary.

6.13.1

Home

menu and parameters

Abb. 6-3:Menu – Home

Fig. 6-3:

Menu –

Home

Parameter

Value

Range of values

Increment

Standard

Home speed

Set the speed of the Home
function

5 – 10000

5

7500

Home offset

Set vertical offset in μm

5 – 20000

5

0