1 application parameters, 2 application – cell transfer, Application parameters – Eppendorf TransferMan4 m User Manual
Page 48: Application – cell transfer
Software
TransferMan
®
4m
English (EN)
48
Application selection
• Select the application
• Store the main application
5.2.1
Application parameters
Predefined softkeys of the different applications.
•
Pos 1
– Save the X, Y and Z values of the capillary position. Switch the position using
the joystick key.
•
Pos 2
– Save the X, Y and Z values of the capillary position. Switch the position using
the joystick key.
•
Pos 3
– Save the X, Y and Z values of the capillary position. Switch the position using
the joystick key.
•
Y off
– Switch off the movement of the capillary in the Y-axis. Prevents moving
sideways during injection.
•
Z-axis Limit
– Set the lower limit for vertical capillary movement.
•
Axial
– Switch on the capillary movement along the mounting bracket.
5.2.2
Application –
Cell transfer
This application is suitable for, e.g., stem cell transfer, polar body biopsy and similar
applications.
Abb. 5-5:Application 1 Cell transfer
Fig. 5-5:
Application 1
Cell transfer
Parameter selection
• Double-click on joystick key – switch position
• Save positions 1, 2 and 3
• Freely program softkey
• Set lower limit (
Z-axis Limit
)