4 deleting the upper limit, 10 horizontal limit, 1 defining the horizontal limit – Eppendorf TransferMan4 m User Manual
Page 61: 2 deleting the horizontal limit, Deleting the upper limit, Horizontal limit 6.10.1, Defining the horizontal limit, Deleting the horizontal limit
Operation
TransferMan
®
4m
English (EN)
61
• The value for the limit (
Lim
) is displayed.
• The capillary cannot be moved any higher.
6.9.4
Deleting the upper limit
1. In the
Installation
menu, select the
Upper limit
submenu.
The
Upper Limit
window appears.
2. Save the position with
Clear
.
• The limit is deleted.
6.10
Horizontal limit
For the X-axis, a limit can be defined for a horizontal injection. This prevents the capillary
from moving through the sample.
6.10.1
Defining the horizontal limit
1. In the
Installation
menu, select the
Angle
submenu.
2. Set the injection angle to 0° and save with
Enter
.
3. Close the menu.
The application screen now displays
X-axis Limit
.
4. Move the capillary to the desired final position for the X-axis.
5. Save the lateral limit with
X-axis Limit
.
• The X-coordinate is marked with
L
.
• The value for the limit (
Limit X
) is displayed.
•
X-axis Limit
is selected.
• The capillary cannot be moved any further to the side.
6.10.2
Deleting the horizontal limit
1. Press
X-axis Limit
.
The limit is deactivated.
2. In the
Installation
menu, select the
Angle
submenu.
3. Reset the injection angle to the operating angle and save with
Enter
.