13.4

Application parameters

Mastercycler nexus gradient, Mastercycler nexus gradient eco

Sample capacity

Universal block, aluminum
96 × 0.1 mL or 0.2 mL PCR tubes
71 × 0.5 mL PCR tubes
one 96-well PCR plate (unskirted, semi-skirted, skirted;
standard and low-profile)

Block temperature control range

4 °C to 99 °C

Ramp rate
(measured at the block)

max. 3°C/s

Block temperature homogeneity
(with the gradient function switched off)

at 35°C
at 90°C

≤ ±0.3°C
≤ ±0.4 °C

Control accuracy
(with the gradient function switched off)

±0.2°C

Gradient range

max. 20°C

Gradient temperature control range

30 °C to 99 °C

Heated lid temperature control range

37 °C to 110 °C

Mastercycler nexus, Mastercycler nexus eco

Sample capacity

Universal block, aluminum
96 × 0.1 mL or 0.2 mL PCR tubes
71 × 0.5 mL PCR tubes
one 96-well PCR plate (unskirted, semi-skirted, skirted;
standard and low-profile)

Block temperature control range

4 °C to 99 °C

Ramp rate
(measured at the block)

max. 3°C/s

Block temperature homogeneity

at 35°C
at 90°C

≤ ±0.3°C
≤ ±0.4 °C

Control accuracy

±0.2°C

Gradient range

Gradient temperature control range

Heated lid temperature control range