1 status view, Status view – Eppendorf Mastercycler nexus User Manual
Page 62
PCR run
Mastercycler
®
nexus
English (EN)
62
7.2.1
Status view
The status view is automatically displayed immediately after a program is started. It shows the progress of
and information on a running program. The currently active step is displayed as a flashing line.
You can exit the status view and return to the navigation tree at any time by pressing the
Exit
softkey, e.g. to
edit another program or start a program on a different cycler. Use the left/right arrow key to access the
status view of other connected cyclers. The program that was already started is not interrupted as a result.
1
User, program name and cycler
2
Remaining run time
3
Start and end time of the program
4
Block and heated lid temperature
5
Device status
6
Cycle number
7
Block target temperature
8
Remaining holding time
9
Information on the current program step
10 Specification of the selected cycler
If you have activated the
TSP
function in the program header (see p. 51), the heated lid is first
brought to the correct temperature before the program run starts. This process is indicated in
the status view by the message
TSP heated lid
.
1
2
3
4
5
6
7
8
9
10