beautypg.com

Eppendorf Mastercycler nexus User Manual

Page 100

background image

Technical data
Mastercycler

®

nexus

English (EN)

100

Mastercycler nexus flat, Mastercycler nexus flat eco

Sample capacity

Flat block, aluminum
4 slides

Block temperature control range

4 °C to 99 °C

Ramp rate
(measured at the block)

max. 3°C/s

Block temperature homogeneity
(with the gradient function switched off)

at 35°C
at 90°C

≤ ±0.3°C
≤ ±0.4 °C

Control accuracy
(with the gradient function switched off)

±0.2°C

Gradient range

max. 20°C

Gradient temperature control range

30 °C to 99 °C

Heated lid temperature control range

37 °C to 110 °C

Mastercycler nexus GX2, Mastercycler nexus GX2e

Sample capacity

Universal block, aluminum
96 × 0.1 mL or 0.2 mL PCR tubes
64 × 0.5 mL PCR tubes
one 64-well PCR plate, one 32-well PCR plate (unskirted,
semi-skirted, skirted; standard and low-profile)

Block temperature control range

4 °C to 99 °C

Ramp rate
(measured at the block)

max. 3°C/s

Block temperature homogeneity
(with the gradient function switched off)

at 35°C
at 90°C

≤ ±0.3°C
≤ ±0.4 °C

Control accuracy
(with the gradient function switched off)

±0.2°C

Thermoblock (64 wells)

Gradient range

max. 12°C

Gradient temperature control range

30 °C to 99 °C

Heated lid temperature control range

37 °C to 110 °C

Thermoblock (32 wells)

no gradient function

Gradient range

Gradient temperature control range

Heated lid temperature control range

37 °C to 110 °C