Safety
Mastercycler
®
nexus
English (EN)
25
CAUTION! Risk of burns on thermoblock, heated lid and reaction vessels.
The thermoblock, heated lid and reaction vessels reach temperatures of more than 50 °C very
quickly.
During gradient operation, the temperature of the heated lid and thermoblock is higher on the
right side than on the left side.
Wait until the temperature of the thermoblock, heated lid and reaction vessels is below 30
°C.
Then the heated lid may be opened.
CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf may
impair the safety, functioning and precision of the device. Eppendorf cannot be held liable or
accept any liability for damage resulting from the use of incorrect or non-recommended
accessories and spare parts, or from the improper use of such equipment.
Only use accessories and original spare parts recommended by Eppendorf.
WARNING! Risk of injury when lifting the device.
If the device is moved using the lid, the lid may break, causing the device to fall.
Lift the device by holding it at the bottom of the housing.
Use both hands to carry the device.
Do not lift the device using the lid.
CAUTION! Risk of injury due to lifting and carrying heavy loads
The device is heavy. Lifting and carrying the device can lead to back injuries.
Transport and lift the device with an adequate number of helpers only.
Use a transport aid to transport the device.
NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool environment to a
warmer environment.
After installing the device, wait for at least 12 h. Only then connect the device to the mains
power supply.