System management
Mastercycler
®
nexus
English (EN)
71
Schedule Settings
: Make
schedule settings
Use the
Schedule Settings
node to configure the schedule view and
selection options.
Mark the
Schedule Settings
node. Press the
Set
softkey.
• Check the
Enable Screensaver
checkbox to use the schedule as a
screensaver. Set the time after which the screensaver should be
activated in the
Wait [s]
field.
• To select the weekdays that should be displayed in the schedule,
check the corresponding boxes for
In schedule day view the following
weekdays are displayed
.
• To select the device types that should be displayed in the schedule,
check the corresponding boxes for
In the schedule editor the following
cyclers are enabled
.
Press the
OK
softkey.
Service Intervals
: Setting
intervals for temperature
verification and self-test
You can determine intervals for a regular temperature verification or
self-test via the
Service Intervals
node.
Mark the
Service Intervals
node. Press the Verification or Self-test
softkey.You can set the intervals In the following display.
• Specification of months (1 – max. 24)
• Specification of programs run (1 – max. 2000)
The following details are shown on the display for each cycler:
• The programs run since the last verification or self-test
• The date of the next check
You can also find this information under the
Block
((see p. 74)) cycler
system functions.
You are informed via an info message before the defined intervals
expire. The info message contains details about the remaining days or
program runs until the interval expires. If email dispatch is activated
((see p. 81)), the administrator receives a corresponding message.
When the intervals expire, a temperature verification or self-test must be
carried out. Alternatively, you can deactivate the interval setting.