10 maintenance, 1 cleaning, 1 cleaning the housing – Eppendorf Mastercycler nexus User Manual
Page 91: Cleaning 10.1.1, Cleaning the housing

Maintenance
Mastercycler
®
nexus
English (EN)
91
10
Maintenance
10.1
Cleaning
10.1.1
Cleaning the housing
Clean the surfaces of the Mastercycler nexus with a damp, lint-free cloth.
If necessary, use a neutral lab cleaner.
Instructions on disinfecting and decontaminating the device can be found separately (see Disinfection/
decontamination on p. 92).
CAUTION! Risk of burns on thermoblock, heated lid and reaction vessels.
The thermoblock, heated lid and reaction vessels reach temperatures of more than 50 °C very
quickly.
During gradient operation, the temperature of the heated lid and thermoblock is higher on the
right side than on the left side.
Wait until the temperature of the thermoblock, heated lid and reaction vessels is below 30
°C.
Then the heated lid may be opened.
NOTICE! Damage from the use of aggressive chemicals.
Do not use any aggressive chemicals on the device or its accessories, such as strong and
weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol.
If the device has been contaminated by aggressive chemicals, clean it immediately using a
mild cleaning agent.
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect the power plug before starting cleaning or
disinfection work.
Do not allow any liquids to penetrate the inside of the housing.
Do not spray clean/spray disinfect the housing.
Only plug the device back in if it is completely dry, both inside and outside.
DANGER! Electric shock.
Switch off the device and disconnect the power plug before starting maintenance or
cleaning work.