To test the time interval accuracy
Testing the time interval accuracy does not check a specification, but does check the
following:
•
125 MHz oscillator
This test verifies that the 125 MHz timing acquisition synchronizing oscillator is
operating within limits.
Equipment Required
Equipment
Critical Specifications
Recommended
Model/Part
Pulse Generator
100 MHz 3.5 ns pulse width, < 600 ps rise time
Agilent 8131A Option 020
Function Generator
Accuracy
≤ (5)(10
−6
)
× frequency
Agilent 3325B Option 002
SMA Cable
Agilent 8120-4948
Adapter
BNC(m)-SMA(f)
Agilent 1250-2015
BNC Test Connector,
6x2
Set up the equipment
1
Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes if you have not already done so.
2
Set up the pulse generator according to the following table.
Pulse Generator Setup
Channel 1
Period
Mode
EXT TRIG
Delay: 0 ps
5 us
TRIG
Slope: Positive
Width: 2.5 us
THRE: 1.0 V
High:
−0.9 V
Low:
−1.7 V
COMP: Disabled
(LED off)
3–54