Agilent Technologies 01664-97005 User Manual
Page 62

Check the setup/hold with single clock edges, multiple clocks
1
Select the logic analyzer setup/hold time.
a
In the logic analyzer Format menu, select Master Clock.
b
Select and activate any two clock edges.
c
Select the Setup/Hold field and select the setup/hold to be tested for all pods. The first
time through this test, use the top combination in the following table.
Setup/Hold Combinations
4.5/0.0 ns
0.0/4.5 ns
2.0/2.5 ns
d
Select Done to exit the setup/hold combinations.
2
Disable the pulse generator channel 2 COMP (with the LED off).
3
Using the Delay mode of the pulse generator channel 1, position the pulses
according to the setup time of the setup/hold combination selected, +0.0 ps or
−100
ps.
a
In the oscilloscope Delta V menu, set the Marker 1 Position to Chan 1, then set
Marker 1 at
−1.3000 V. Set the Marker 2 Position to Chan 2, then set Marker 2 at
−1.3000 V.
b
In the oscilloscope Delta T menu, select Start on Pos Edge 1. Select Stop on Pos
Edge 1.
Testing Performance
To test the multiple-clock, multiple-edge, state acquisition
3–40