TABLE 10-4. ConfigCmd Values for Setting and Retrieving Settings
ConfigCmd
Variable
Setting Description
(some settings list possible values for the DestSource
variable)
Set
Retrieve
0
100
Bandwidth: 5 = 5 Hz, 10 = 10 Hz, 12 = 12.5 Hz, 20 =
20 Hz, 25 = 25 Hz
1
101
Unprompted output: 10 = 10 Hz, 25 = 25 Hz, 50 = 50
Hz
2
102
Pressure sensor: 0 = EC100 basic, 1 = user-supplied, 2
= EC100 enhanced, 3 = none (use fixed value)
3
103
Differential pressure: 0 = disable, 1 = enable
4
104
Fixed pressure value
5
105
Pressure offset
6
106
Pressure gain
7
107
Temperature sensor:
0 = IRGASON/EC150 temperature probe
1 = EC155 sample cell thermistor
2 = EC155 sample cell thermocouple
3 = None (use fixed value)
4 = Auto-select
8
108
Fixed temperature value
9
109
Unprompted output mode: 0 = Disable, 1 =USB, 2 =
RS-485
10
110
RS-485 baud rate
11
111
Zero-and-span control: 0 = inactive, 1 = zero, 2 = span
CO
2
, 3 = span H
2
O (See Section 10.2.1, ConfigCmd 11
Zero-and-span Control)
12
112
CO
2
span concentration
13
113
H
2
O span dew point temperature
14
114
CO
2
zero
15
115
CO
2
span
16
116
H
2
O zero
17
117
H
2
O span
18 or
218
118
Heater voltage (0 to 4.5375V, −1 = Off) (See Section
10.2.2, ConfigCmd 18 Heater Voltage)
19
119
Reserved
20
120
Analog output enable: 0 = disable, 1 = enable
21
121
Power down: 0 = Gas head on, 1 = Gas head off
99
N/A
Save settings to EEPROM memory
DestSource
is a variable that will contain the value to read when getting a
setting, or that will contain the value to send when writing a setting to the
IRGASON.