Campbell Scientific IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer User Manual

Page 48

background image

IRGASON

®

Integrated CO

2

/H

2

O Open-Path Gas Analyzer and 3D Sonic Anemometer

TABLE 10-4. ConfigCmd Values for Setting and Retrieving Settings

ConfigCmd

Variable

Setting Description

(some settings list possible values for the DestSource

variable)

Set

Retrieve

0

100

Bandwidth: 5 = 5 Hz, 10 = 10 Hz, 12 = 12.5 Hz, 20 =

20 Hz, 25 = 25 Hz

1

101

Unprompted output: 10 = 10 Hz, 25 = 25 Hz, 50 = 50

Hz

2

102

Pressure sensor: 0 = EC100 basic, 1 = user-supplied, 2

= EC100 enhanced, 3 = none (use fixed value)

3

103

Differential pressure: 0 = disable, 1 = enable

4

104

Fixed pressure value

5

105

Pressure offset

6

106

Pressure gain

7

107

Temperature sensor:

0 = IRGASON/EC150 temperature probe

1 = EC155 sample cell thermistor

2 = EC155 sample cell thermocouple

3 = None (use fixed value)
4 = Auto-select

8

108

Fixed temperature value

9

109

Unprompted output mode: 0 = Disable, 1 =USB, 2 =

RS-485

10

110

RS-485 baud rate

11

111

Zero-and-span control: 0 = inactive, 1 = zero, 2 = span

CO

2

, 3 = span H

2

O (See Section 10.2.1, ConfigCmd 11

Zero-and-span Control)

12

112

CO

2

span concentration

13

113

H

2

O span dew point temperature

14

114

CO

2

zero

15

115

CO

2

span

16

116

H

2

O zero

17

117

H

2

O span

18 or

218

118

Heater voltage (0 to 4.5375V, −1 = Off) (See Section

10.2.2, ConfigCmd 18 Heater Voltage)

19

119

Reserved

20

120

Analog output enable: 0 = disable, 1 = enable

21

121

Power down: 0 = Gas head on, 1 = Gas head off

99

N/A

Save settings to EEPROM memory

DestSource

is a variable that will contain the value to read when getting a

setting, or that will contain the value to send when writing a setting to the

IRGASON.

38