beautypg.com

Maintenance, 1 routine maintenance, 2 wick maintenance – Campbell Scientific IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer User Manual

Page 35: 1 sonic wicks, Routine maintenance, Wick maintenance, Sonic wicks, 2. multipliers and offsets for analog outputs

background image

IRGASON

®

Integrated CO

2

/H

2

O Open-Path Gas Analyzer and 3D Sonic Anemometer

TABLE 8-2. Multipliers and Offsets for Analog Outputs

Density

(mg m

-3

)

Voltage Output Multiplier

(mg m

-3

V

-1

)

Offset

(mg m

-3

)

CO

2

386.32

-102.59

H

2

O 8.65

-2.26

9. Maintenance

IRGASON operation requires six maintenance tasks:

routine maintenance

wick maintenance

analyzer window cleaning

zero and span

replacing the analyzer desiccant / scrubber bottles

factory recalibration

9.1 Routine Maintenance

Periodically do the following:

Check the humidity indicator card in the EC100 enclosure. If the highest

dot has turned pink, replace or recharge the desiccant bags. Replacement

desiccant may be purchased (pn 6714) or old desiccant packs may be

recharged by heating in an oven. See the manual ENC10/12, ENC12/14,

ENC14/16, ENC16/18, available at

www.campbellsci.com

, for more details

on recharging desiccant bags.

Make sure the LED Status Lights on the EC100 panel are green. If not,

verify that all connections are secure and that the instruments are powered.

Also check the individual diagnostic bits for the specific fault. See

TABLE 10-2, Bits in the Sonic Diagnostic Flag and TABLE 10-3, Bits in

the Gas Diagnostic Flag.

9.2 Wick Maintenance

9.2.1 Sonic Wicks

The IRGASON cannot measure wind when water droplets completely obscure

the face of the sonic transducers. If the site experiences frequent rainfall, using

wicks on the sonic transducers may be desirable. The wicks help remove

water. Under some conditions, however, droplets may build up and remain on

the transducer faces. As soon as the water droplets evaporate or are removed,

the IRGASON will again measure wind. Remove the water droplets by

dabbing a cotton swab or tissue on the faces of the transducers. The user must

use discretion to determine if wicks are necessary. If wicks are not used,

gently remove the wicks from the sonic transducers, taking care not to damage

or peal the matching layer (rubber tips) from the brass housing of the

transducers.

Remove the wicks during the winter. The wicks will accumulate snow or

freezing rain, resulting in blockage of the sonic signals. The CSAT3 Sonic

25