7.1 Factory Defaults

TABLE 7-1 shows the default value for each setting described in the following

sections.

TABLE 7-1. Factory Default Settings

Bandwidth

20 Hz

SDM Address

1

Unprompted Output

disabled

RS-485 baud rate

115200 bps

Unprompted Output Rate 10 Hz
Analog Output

disabled

ECMon Update Rate

10 Hz

Temperature Sensor

Auto-Select (IRGASON Temp Sensor)

Pressure Sensor

EC100 Basic or EC100 Enhanced (depending on

order)

Pressure Differential

Enable

Auto-Select (Disabled for IRGASON)

Heater Control

Disabled

7.2 Bandwidth

The EC100 has a user-selectable low-pass filter to select the bandwidth (5, 10,

12.5, 20, or 25 Hz). Setting the bandwidth to a lower value will reduce noise.

However, it must be set high enough to retain the high-frequency fluctuations

in CO

2

and H

2

O, or else the high frequency contributions to the flux will be

lost. The factory default bandwidth of the EC100 is 20 Hz, which is sufficient

for most flux applications. Lower bandwidth settings may be used for higher

measurement heights, which inherently have lower-frequency content. Refer

to Appendix A, Filter Bandwidth and Time Delay, for more information on the

digital filter options.

If a spectral analysis is being done to evaluate the experimental setup, the

bandwidth should be set to the Nyquist frequency, which is half the datalogger

sample rate (for SDM output) or half the unprompted output rate (for USB and

RS-485 output). This ensures that the data will not be under-sampled and that

higher-frequency variations will not be aliased to lower frequencies. Note that

if too small a bandwidth is selected, high-frequency fluxes may be under-

measured.

7.3 SDM Address

This parameter must be set to use SDM output from the EC100. See Section

8.1, SDM Output, for details on using SDM output.

Each SDM device on the SDM bus must have a unique address. The

IRGASON has a factory default SDM address of 1, but it may be changed to

16