beautypg.com

Campbell Scientific IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer User Manual

Page 40

background image

IRGASON

®

Integrated CO

2

/H

2

O Open-Path Gas Analyzer and 3D Sonic Anemometer

FIGURE 9-4. The zero-and-span shroud mounted on the zero-and-

span stand. The top tube is the gas intake, the black cable is the

temperature thermistor, and the short bottom tube is the gas

exhaust.

4. Disconnect the IRGASON temperature sensor from the EC100 and

connect the shroud temperature sensor in its place.

5. Connect the EC100 to a PC with the EC100 USB cable (pn 26563).

6. Resume power to the EC100. Wait for all the

Gas and Power LED status

lights on the EC100 panel to turn green.

7. Launch ECMon, select the appropriate USB port, and click

Connect. The

main screen should now be reporting real-time CO

2

and H

2

O

concentrations. Click

Zero/Span. A graph will appear in the lower half

of the zero-and-span window showing measured CO

2

and H

2

O

concentrations (see FIGURE 9-5).

8. Connect a gas tank of known CO

2

concentration to a pressure regulator,

then to a flow controller, and finally to the intake of the shroud.

Optimally, the concentration of span CO

2

should be near the concentration

of CO

2

being measured in the field. Begin with both the pressure regulator

and flow controller turned off. Then use the pressure regulator to slowly

increase pressure to the recommended setting for the flow controller. Next

set the flow between 0.4 and 0.6 l/min. Watch the ECMon zero-and-span

graph for the CO

2

measurement readings to stabilize (5 to 10 minutes).

Once stable, write down the reported CO

2

concentration.

Intake

Thermistor Cable

Exhaust

30