2 output signals, 3 physical description, Output signals – Campbell Scientific IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer User Manual
Page 15: Physical description

IRGASON
®
Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
5.2 Output Signals
Features
•
EC100 electronics outputs data using:
o CS SDM
o RS-485
o USB
o Analog out
Digital
SDM (Synchronous Device for Measurement)
1
Data type:
FLOAT
RS-485
Data type:
ASCII
Output Rate:
5 to 50 Hz
Baud rate:
1200 to 230400 bps
USB
Data type:
ASCII
Output rate:
10, 25 or 50 Hz
Analog (two outputs for CO
2
and H
2
O densities)
Voltage range:
0 mV to 5000 mV
Resolution:
76 µV (16 bit)
Update rate:
100 Hz
Accuracy (at 25
o
C):
±3 mV
CO
2
density equation:
mg·m
–3
= 0.38632· (mV
out
) – 102.59
Full scale range:
–103 to 1829 mg·m
–3
H
2
O density equation:
g·m
–3
= 0.00865· (mV
out
) – 2.26
Full scale range:
–2 to 41 g·m
–3
1
Synchronous Device for Measurement. A Campbell Scientific, Inc. proprietary serial interface
for datalogger to peripheral and sensor communication. See Section 8.1, SDM Output, for details.
2
user selectable
5.3 Physical Description
Optical path length:
15.37 cm (6.05 in)
Sonic path length:
10.0 cm
Sonic path angle from vertical: 30°
Spatial separation between gas analyzer and sonic volumes: 0 cm
Dimensions
IRGASON
Length:
51.5 cm (20.3 in)
Height:
31.1 cm (12.2 in)
Cable Length:
3 m (9.8 ft)
EC100 Enclosure:
24.1 x 35.6 x 14 cm (9.5 x 14.0 x 5.5 in)
Weight
IRGASON with Cable:
2.8 kg (6.1 lbs)
EC100 with Enclosure:
3.2 kg (7 lbs)
5