beautypg.com

2 output signals, 3 physical description, Output signals – Campbell Scientific IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer User Manual

Page 15: Physical description

background image

IRGASON

®

Integrated CO

2

/H

2

O Open-Path Gas Analyzer and 3D Sonic Anemometer

5.2 Output Signals

Features

EC100 electronics outputs data using:

o CS SDM

o RS-485

o USB

o Analog out

Digital

SDM (Synchronous Device for Measurement)

1

Data type:

FLOAT

RS-485

Data type:

ASCII

Output Rate:

5 to 50 Hz

Baud rate:

1200 to 230400 bps

USB

Data type:

ASCII

Output rate:

10, 25 or 50 Hz

Analog (two outputs for CO

2

and H

2

O densities)

Voltage range:

0 mV to 5000 mV

Resolution:

76 µV (16 bit)

Update rate:

100 Hz

Accuracy (at 25

o

C):

±3 mV

CO

2

density equation:

mg·m

–3

= 0.38632· (mV

out

) – 102.59

Full scale range:

–103 to 1829 mg·m

–3

H

2

O density equation:

g·m

–3

= 0.00865· (mV

out

) – 2.26

Full scale range:

–2 to 41 g·m

–3

1

Synchronous Device for Measurement. A Campbell Scientific, Inc. proprietary serial interface

for datalogger to peripheral and sensor communication. See Section 8.1, SDM Output, for details.

2

user selectable

5.3 Physical Description

Optical path length:

15.37 cm (6.05 in)

Sonic path length:

10.0 cm

Sonic path angle from vertical: 30°
Spatial separation between gas analyzer and sonic volumes: 0 cm
Dimensions

IRGASON

Length:

51.5 cm (20.3 in)

Height:

31.1 cm (12.2 in)

Cable Length:

3 m (9.8 ft)

EC100 Enclosure:

24.1 x 35.6 x 14 cm (9.5 x 14.0 x 5.5 in)

Weight

IRGASON with Cable:

2.8 kg (6.1 lbs)

EC100 with Enclosure:

3.2 kg (7 lbs)

5