9.3 Gas-Analyzer Window Cleaning

The windows of the analyzer should be cleaned if the signal strength for CO

2

or H

2

O drops below 0.7 or 70% of the original value. These values may be

monitored in the output data, or they can be viewed with ECMon. To clean the

windows, use isopropyl alcohol (2-propanol / isopropanol) and a cotton swab

or a non-scratching tissue or cloth. Verify that following cleaning the signal

strengths are restored to values close to 1.0.

9.4 Zero and Span

As is the case with all optical instruments, the IRGASON’s gas analyzer

measurements may drift slightly with exposure to natural elements. Therefore,

routine maintenance requires a periodic zero-and-span procedure. The

frequency with which this should be done is heavily dependent on the

environmental conditions in which the instrument is making measurements.

Campbell Scientific recommends that a new instrument, or an instrument that

has been moved to a new location, should have a monthly zero and span

procedure performed. Additionally, it is recommended to always zero and

span the analyzer after cleaning the windows (see Section 9.3, Gas-Analyzer

Window Cleaning). After the first several zero-and-span procedures, the rate of

drift in gain and offset (explained later in this section) will be better understood

for the existing environmental conditions. A regimented protocol for zero and

span of the instrument initially, will allow the user to assess the appropriate

time interval between zero and span procedures. However, if the IRGASON is

being used at a site with large seasonal changes in ambient conditions, the

zero-and-span procedure should be done at least seasonally.

In situations in which spanning the instrument is extremely difficult (for

example, in below freezing temperature conditions or in very remote sites),

performing a monthly zero will help mitigate the drift in the analyzer’s

measurements.