beautypg.com

1. an example of usb data output in terminal mode, 1. usb and rs-485 output elements – Campbell Scientific IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer User Manual

Page 33

background image

IRGASON

®

Integrated CO

2

/H

2

O Open-Path Gas Analyzer and 3D Sonic Anemometer

To use USB or RS-485 output, connect a USB or RS-485 cable from the

EC100 to the receiving device (see Section 6.3, Wiring and Connections), and

configure the settings (see Section 7, Settings).

The

Unprompted Output setting must be set to USB or RS-485.

If

RS-485 is selected, the RS485 Baud Rate must be set.

The

Unprompted Output Rate must be set to the desired output

rate.

All output data are formatted as comma delimited ASCII. Each record

terminates with a carriage return and line feed. TABLE 8-1 lists the elements

in each output array, and FIGURE 8-1 shows an example USB data feed in

terminal mode.

TABLE 8-1. USB and RS-485 Output Elements

Data Element Description

Units/comments

1

U

x

m/s

2

U

y

m/s

3

U

z

m/s

4

Sonic temperature

°C

5

Sonic diagnostic flag

6

CO

2

density

mg/m

3

7

H

2

O density

g/m

3

8

Gas diagnostic flag

9

Air temperature

°C

10

Air pressure

kPa

11

CO

2

signal strength

Nominally 0.0 to 1.0

12

H

2

O signal strength

Nominally 0.0 to 1.0

13

Not used

14

Counter

Arbitrary

15

Signature

Arbitrary in hexadecimal

FIGURE 8-1. An example of USB data output in terminal mode

23