IRGASON

®

Integrated CO

2

/H

2

O Open-Path Gas Analyzer and 3D Sonic Anemometer

Sonic accuracy

5

Offset error

u

x

, u

y

:

< ±8 cm·s

–1

u

z

:

< 4 cm·s

–1

Gain error

Wind vector ±5° horizontal: < ±2% of reading

Wind vector ±10° horizontal: < ±3% of reading

Wind vector ±20° horizontal: < ±6% of reading

Sonic reporting range

Full scale wind:

±65.553 m/s

Sonic temperature:

–50° to +60°C

Auxiliary sensors

6

Barometer

Basic barometer

Accuracy

–30 to 0°C:

±3.7 kPa at –30°C, falling linearly to ±1.5

kPa at 0

o

C

0° to 50°C:

±1.5 kPa

Measurement rate: 10 Hz

Optional enhanced barometer:

Manufacturer:

Vaisala

Model:

PTB110

Accuracy

–30 to 0°C:

±0.15 kPa

Measurement rate: 1 Hz

IRGASON temperature sensor:

Manufacturer: BetaTherm

Model:

100K6A1A Thermistor

Accuracy:

±0.15

o

C (–30

o

to 50

o

C)

1

noise rms, assumes:

o 25°C

o 85 kPa

o 14 g·m

–3

H

2

O

o 597 mg·m

–3

CO

2

o 25 Hz bandwidth.

2

user selectable

3

–30° to 50°C

4

noise rms

5

assumes:

o –30° to +50°C

o wind speed <30 m·s

–1

o azimuth angles between ±170°

6

refer to manufacturer’s product brochure or manual for details