3. Connect the IRGASON temperature probe. Unscrew the temperature

connector cover, which is found on the base of the EC100 enclosure (see

FIGURE 6-6). Insert the 3-prong probe connector into the female

connector on the enclosure and screw it firmly in place. The IRGASON

temperature probe cable is approximately 3 meters in length.

4. Ground the EC100 by attaching a thick wire (e.g., 12 AWG) to the

grounding lug found on the base of the EC100 enclosure. The other end of

the wire should be connected to earth (chassis) ground using a grounding

rod. For more details on grounding, see the CR3000 datalogger manual,

grounding section.

Grounding electronics such as the EC100 is critical in a

measurement system. Proper grounding to earth (chassis)

will ensure maximum ESD (electrostatic discharge)

protection and improve measurement accuracy.

Do not connect or disconnect the gas analyzer or sonic

connectors while the EC100 is powered.

5. Connect a communications signal cable to the EC100. Loosen the nut on

one of the cable entry seals (Cable 1 or Cable 2) on the base of the EC100

enclosure, remove the plastic plug (plug may be stored in mesh pocket of

enclosure), insert the cable, and retighten the nut by hand. Refer to the

sections below on SDM, USB, RS-485, and Analog communications for

information on required signal cable types and connections to the EC100

panel.

o SDM communications: Use cable CABLE4CBL-L (pn 21972).

“L” denotes the length of the cable, which is customer-specified

at time of order. TABLE 6-1

details which color of wire in the

cable should be connected to each terminal found on the SDM

connector of the EC100 panel.

TABLE 6-1. EC100 SDM Output to a Campbell Scientific CR1000,

CR3000, or CR5000 Datalogger

EC100 Channel

Description

Color

SDM-C1

SDM Data

Green

SDM-C2

SDM Clock

White

SDM-C3

SDM Enable

Red (or Brown)

G

Digital Ground

Black

G

Shield

Clear

o USB communications: Use the EC100 USB cable (pn 26563) to

connect a PC to the base of the EC100 enclosure.

o RS-485 communications: use cable CABLE3TP-L (pn 26987)

for lengths less than 500 ft. The connector on the EC100 panel

labeled RS-485 displays which terminals are for receiving and

transmitting.