1. output modes for ec100 instruction, 2. bits in the sonic diagnostic flag – Campbell Scientific IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer User Manual

Page 45

background image

IRGASON

®

Integrated CO

2

/H

2

O Open-Path Gas Analyzer and 3D Sonic Anemometer

TABLE 10-1. Output Modes for EC100 Instruction

Output

Mode

Data

Field

Description

Units

0, 1, 2,

1

Ux

m/s

2

Uy

m/s

3

Uz

m/s

4

Sonic Temperature

ºC

5

Sonic Diagnostic Flag

6

CO

2

Density

mg/m

3

7

H

2

O Density

g/m

3

8

Gas Diagnostic Flag

1, 2

9

Air Temperature

ºC

10

Air Pressure

kPa

11

CO

2

Signal Strength

nominally 0.0 ≤ strength

≤1.0

12

H

2

O Signal Strength

nominally 0.0 ≤ strength

≤1.0

2

13

Sample Cell Pressure

Differential

kPa

As shown in TABLE 10-1, all output modes give two diagnostic values:

Sonic

Diagnostic Flag and Gas Diagnostic Flag. The values contain a bit field with

each bit representing a monitored condition. When a certain condition is

detected, the corresponding bit is set. The value remains set until the event that

caused the condition is no longer present. TABLE 10-2 and TABLE 10-3

describe the bits in the sonic diagnostic flag and the gas diagnostic flag,

respectively.

TABLE 10-2. Bits in the Sonic Diagnostic Flag

bit

hex

value

decimal

Name

Function

0

0x1

1

Low Amp

Amplitude is too low

1

0x2

2

High Amp

Amplitude is too high

2

0x4

4

Tracking

Poor signal lock

3

0x8

8

Hi 3 Axis DC

Delta temperature exceeds

limits

4

0x10

16

Acquiring

Acquiring ultrasonic

signals

5

0x20

32

Cal Mem Err

Sonic head calibration

signature error