1. output modes for ec100 instruction, 2. bits in the sonic diagnostic flag – Campbell Scientific IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer User Manual
Page 45

IRGASON
®
Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
TABLE 10-1. Output Modes for EC100 Instruction
Output
Mode
Data
Field
Description
Units
0, 1, 2,
1
Ux
m/s
2
Uy
m/s
3
Uz
m/s
4
Sonic Temperature
ºC
5
Sonic Diagnostic Flag
6
CO
2
Density
mg/m
3
7
H
2
O Density
g/m
3
8
Gas Diagnostic Flag
1, 2
9
Air Temperature
ºC
10
Air Pressure
kPa
11
CO
2
Signal Strength
nominally 0.0 ≤ strength
≤1.0
12
H
2
O Signal Strength
nominally 0.0 ≤ strength
≤1.0
2
13
Sample Cell Pressure
Differential
kPa
As shown in TABLE 10-1, all output modes give two diagnostic values:
Sonic
Diagnostic Flag and Gas Diagnostic Flag. The values contain a bit field with
each bit representing a monitored condition. When a certain condition is
detected, the corresponding bit is set. The value remains set until the event that
caused the condition is no longer present. TABLE 10-2 and TABLE 10-3
describe the bits in the sonic diagnostic flag and the gas diagnostic flag,
respectively.
TABLE 10-2. Bits in the Sonic Diagnostic Flag
bit
hex
value
decimal
Name
Function
0
0x1
1
Low Amp
Amplitude is too low
1
0x2
2
High Amp
Amplitude is too high
2
0x4
4
Tracking
Poor signal lock
3
0x8
8
Hi 3 Axis DC
Delta temperature exceeds
limits
4
0x10
16
Acquiring
Acquiring ultrasonic
signals
5
0x20
32
Cal Mem Err
Sonic head calibration
signature error
35