beautypg.com

5. ecmon zero-and-span window – Campbell Scientific IRGASON Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer User Manual

Page 41

background image

IRGASON

®

Integrated CO

2

/H

2

O Open-Path Gas Analyzer and 3D Sonic Anemometer

Use mixtures of CO

2

and ambient air for the CO

2

span gas. The

use of reference CO

2

gas mixtures in pure nitrogen will lead to

errors due to a carrier gas effect on pressure-broadening of the CO

2

absorption lines since oxygen gas has a smaller line-broadening

coefficient than nitrogen.

FIGURE 9-5. ECMon zero-and-span window

9. Remove the CO

2

span gas from the inlet of the shroud, and replace it with

H

2

O span gas from a dew-point generator or another standard reference.

Because water molecules can adsorb to inside of the tubing and the shroud,

it may take many minutes for the H

2

O concentration to stabilize. If

desired, increase the flow rate for the first several minutes to more quickly

stabilize the system before returning it to between 0.4 and 0.6 l/min to

make the H

2

O measurement. Write down the reported H

2

O concentration.

10. Remove the H

2

O span gas, and connect a zero air source (no CO

2

or H

2

O)

to the inlet tube of the shroud. As described in step 8, use a pressure

regulator and flow controller so that zero air flows through the shroud

between 0.4 and 0.6 l/min. Wait for the measurement readings to stabilize

(this may require several minutes) and write down the reported values for

CO

2

and H

2

O concentrations. If the readings remain erratic, ensure that

flow of the zero air is sufficient and the shroud is correctly seated on the

snouts.

NOTE

31