5.4 Power Requirements During Power-up and Operation

Power requirement

Source:

10 to 16 Vdc

Load

at 25

o

C:

420 mA

at 25

o

C in power-down mode

1

: 250 mA

1

sonic anemometer fully powered and gas analyzer in stand-by

6. Installation

6.1 Orientation

During operation, the IRGASON should be leveled and pointed into the

prevailing wind to minimize flow distortions from the instrument arms and

other supporting structures. Leveling can be done using the leveling bubble on

top of the instrument block. In the proper orientation, the product label will

read right-side-up, with the upper arm (source) of the gas analyzer directly

above the lower arm (detector).

6.2 Mounting

The IRGASON is supplied with mounting hardware to attach it to the end of a

horizontal pipe with a diameter of 1.31 in., such as the Campbell Scientific

CM202, CM204, or CM206 crossarm. The following steps describe the

normal mounting procedure. Refer to FIGURE 6-1.

1. Secure the chosen crossarm to a tripod or other vertical structure using a

CM210 Crossarm-to-Pole Mounting Kit (pn 17767).

2. Point the horizontal arm into the direction of the prevailing wind and

tighten the nuts and bolts.

Do not carry the IRGASON by the arms or the strut

between the arms. Always hold it by the block, where

the upper and lower arms connect.

3. Bolt the IRGASON mounting bracket (pn 27496) to the CM250 Leveling

Mount (pn 26559) (see FIGURE 6-1).

4. Install the IRGASON to the mounting bracket by inserting the mounting

bolt and pin into the corresponding holes. Tighten the bolt. Make sure the

analyzer is vertically oriented so the label is right-side-up and the upper

arm (source) is directly above the lower arm (detector).

5. Mount the IRGASON

assembly by fitting the boom adapter over the end

of the crossarm. Tighten the set screws on the boom adapter with a 3/16

inch hex socket head wrench.

WARNING